Clamp Meter
Service Times: Sunday - Friday , 09am and 06:00pm
Call us at 021-29478974 , 021-22997129 , Mobile/WA : 0811 1388 063 , Email : sales@ptbestinstrument.com
MODEL 2002R
 
Selection Guide Instruction Manual
 

Can measure large AC current up to 2000A.
Peak hold function.
55mm-dia large tear drop shaped jaws.

TRUE RMS
Ø55 MAX AC2000A ACV DCV
Ω PEAK 10ms OUT PUT Continuity
2002R
CE
 
AC A 400A (0 - 400A)
|±1.5%rdg±3dgt [45 - 65Hz]
|±2.5%rdg±3dgt [40Hz - 1kHz]
2000A (0 - 1500A)
|±2%rdg±5dgt [45 - 65Hz]
|±3%rdg±5dgt [40Hz - 1kHz]
2000A (1501 - 2000A)
|±4%rdg [50/60Hz]
AC V 40/400/750V 
|±1%rdg±2dgt [45 - 65Hz]
|±1.5%rdg±3dgt [40Hz - 1kHz]
DC V 40/400/1000V ±1%rdg±2dgt
Ω 400Ω/4k/40k/400kΩ|±1.5%rdg±2dgt
Continuity buzzer buzzer sounds below 50±35Ω
Output
Recorder : DC400mV against AC400A
  DC200mV against AC2000A
Conductor Size Ø55mm max.
Safety Standard
IEC 61010-1 CAT III 600V
    CAT II 1000V
IEC 61010-2-031
IEC 61010-2-032
Frequency Response 40Hz - 1kHz
Withstand Voltage 5550V AC for 1 minute
Power Source R6 (AA) (1.5V) × 2
Dimensions 247(L) × 105(W) × 49(D)mm
Weight 470g approx.
Accessories 7107A (Test leads) 
8201
 (Output Plug)
9094 (Carrying Case) 
R6 (AA) × 2
Instruction Manual
Optional 8008 (Multi-Tran)
7256 (Output cord)

 

     

 

STORE

  • Plaza Kenari Mas , Lt. LG Blok B121 , Jalan Kramat Raya No. 101 , Paseban, Senen , JakPus 10440 , Phone : (021)-29478925

  • LTC Glodok Lt. 2 Blok C3 No. 3-5 , Jl. Hayam Wuruk No.127, Mangga Besar, Tamansari, Kota, Jakarta Barat 11180 , Phone : (021)22685705

OFFICE

Ruko Concordia SRB 26, Kota Wisata Cibubur Bogor 16968 Indonesia

Phone : (0 21)-29478974  

Mobile/WA : 0811 1388 063

 Email : sales@ptbestinstrument.com